Near Ultra-violet (NUV) Objective Lens( PAL-100-NUV-LC11-A )

PAL-100-NUV-LC11-A

This objective lens can be used for laser machining using pulsed laser of THG (355nm) YAG laser. Its glass-thickness- compensation optical design makes it possible to realize an ideal beam spot size and quality even if it was processed thorough a cover glass.
◦Two kinds objective lenses are available. They are designed to correct aberration depending on the thickness of cover glass. (t= 0.7 mm and 1.1 mm)
◦High resolution type (NA=0.65) is also available.
◦With its long working infinity correction function; this objective lens can be used for a laser system and coaxial observation.
◦This objective lens can be used with a pulse laser of visible light (532nm).
◦Laser Damage Threshold(reference): 0.05 J/cm2 (355nm), 0.1J/cm2 (532nm) (Laser pulse width: 10ns, repetition frequency: 20Hz)
More Information
Name Near Ultra-violet (NUV) Objective Lens( PAL-100-NUV-LC11-A )
Weight 0.5000kgs
Standard Coatings Available No
Guide
  • Available fixed objective lens holder (LHO-26).
  • WEB Reference Catalog Code/W4024
  • When the objective lens is fixed to a 2 axis holder, please consult our Sales Division.
  • For laser processing, we offer a dichroic block (DIMC) and for laser unit with coaxial illumination and observation (OUCI-2).
  • WEB Reference Catalog Code/W2041, WEB Reference Catalog Code/W2012
Attention
  • When an objective lens is used in laser processing, use the diameter of the incident beam to extend to a size of half the pupil diameter (1/e2). A small light spot cannot be achieved when the incident beam is too narrow. Please note if there is a laser energy density increase, there will be a high possibility of damage to the objective lens.
  • When the thickness of cover glass is not same as the specified, designed specifications may not be achieved due to aberration.
  • Magnification is the value when using the imaging lens f=200mm. When used in a microscope lens barrel from other manufacturers there may be different magnifications. The actual magnification should be calculated from the ratio of the focal length of the objective lens and the focal length of the imaging lens to verify the focal length of the imaging lens barrel to be used.
Focal length 2mm
Magnification 100×
Numerical Aperture 0.57
Working Distance WD 11.13mm
Item name LCD MPlanApo NUV 100x (t0.7)
Focal length f 2mm
Resolution (λ = 550nm) 0.48μm
Focal depth (λ = 550nm) ±0.8μm
Pupil diameter φ2.3mm
Cover glass material Synthetic Quartz
Real field of View (Imaging device 1/2-inch) 0.05×0.06mm
In stock
SKU
PAL-100-NUV-LC11-A
$7,365.00

Delivery in : 6-12 weeks

This objective lens can be used for laser machining using pulsed laser of THG (355nm) YAG laser. Its glass-thickness- compensation optical design makes it possible to realize an ideal beam spot size and quality even if it was processed thorough a cover glass.