(Reference) Durability Data of Manual Ball Guide Stages
Linear Stage
Reciprocate the stage with a load placed vertically on its upper surface,
and measure the state change at that time.
Measurement Sample | TSD-601 |
---|---|
Load | 100N |
Procedure | 13 mm × Both ways/sec (continuous operation) |
Operation Cycles (Distance) | 10 million times (260km) |
TSD series
After 10 million cycles of operation, the guide performance is stable without significant change in the stiffness, straightness, pitching, yawing, and running parallelism of the stage guide.
![TSD series](/pub/media/tutorial/manual-positioners/mp08_img01.jpg)
![Moment Stiffness](/pub/media/tutorial/manual-positioners/mp08_img02.jpg)
![Straightness](/pub/media/tutorial/manual-positioners/mp08_img03.jpg)
![Pitch / Yaw](/pub/media/tutorial/manual-positioners/mp08_img04.jpg)
![Running Parallelism](/pub/media/tutorial/manual-positioners/mp08_img05.jpg)
Goniometer Stage
Reciprocate the stage with a load placed vertically on its upper surface, and measure the state change at that time.
Measurement Sample | GOHT-60A85B |
---|---|
Load | 50N |
Procedure | ±15° × One way/sec (continuous operation) |
Operation Cycles (Distance) | 600,000 times (30 km) |
GOHT series
After 600,000 cycles of operation, the guide performance is stable without significant change in the stiffness and rotation center displacement of the stage guide.
![GOHT series](/pub/media/tutorial/manual-positioners/mp08_img06.jpg)
![Moment Stiffness](/pub/media/tutorial/manual-positioners/mp08_img07.jpg)
![Rotation Center Displacement](/pub/media/tutorial/manual-positioners/mp08_img08.jpg)
Rotation Stage
Reciprocate the stage with a load placed vertically on its upper surface, and measure the state change at that time.
Measurement Sample | KSPT-606MH |
---|---|
Load | 100N |
Procedure | ±10° × Both ways/sec (continuous operation) |
Operation Cycles (Distance) | 100,000 times (1km) |
KSPT series
After 100,000 cycles of operation, the guide performance is stable without significant change in the stiffness, wobble, and parallelism of the stage guide.
![KSPT series](/pub/media/tutorial/manual-positioners/mp08_img09.jpg)
![Moment Stiffness](/pub/media/tutorial/manual-positioners/mp08_img10.jpg)
![Wobble / Parallelism](/pub/media/tutorial/manual-positioners/mp08_img11.jpg)